Constant temperature and humidity test box does not have humidity due to humidity test
The humidity in the constant temperature and humidity test box appears to be 100% with the following possibilities. Simply remove them one by one.
A. Check whether the temperature sensor and the humidity sensor are installed
B. Check whether the wet bulb gauze is hung on the humidity sensor. The sensor near the sink is a humidity sensor. The gauze should be hung on the humidity sensor and wrap the temperature sensor part of the sensor.
C. Check the gauze after the position of the gauze is correct, check whether it can normally absorb water, whether the gauze itself is wet, whether the bottom of the gauze is immersed in the water tank, and whether there is water in the water tank. The above points are eliminated one by one.
Immediately solve the problem, solve the problem still can not solve, continue to look down, for gauze does not absorb water, change the new gauze.
D, if more than three are not solve the problem, after examination are normal, check the heat chamber controller or sensor, which is rare, the need for a professional home inspection, whether the problem is the sensor and the controller itself .